Purpose
Ultrasound transducers are prone to damages that cause deterioration of image quality and Doppler accuracy 1.
There are only few studies that have investigated the proportion defective transducers in daily clinical use 2-4,
numbers extracted from these studies indicate the proportion of defective transducers to lie in the range of 17 – 40 % of all the transducers in daily use.
Traditionally ultrasound quality control (QC) protocols have utilized tissue-mimicking phantoms 5-8.
Although there has been raised criticism towards the use of phantoms in routine...
Methods and Materials
Two identical 12 MHz linear transducers (Toshiba PLT-1204AT) attached to Toshiba Aplio 50 (Toshiba Medical Systems) ultrasound scanner were used.
One of the transducers was fully functional whereas the other had suffered extensive element damage visible in clinical images as well as in images taken with QC phantom (CIRS model 040,
Computerized Imaging Reference Systems,
Norfolk,
VA,
U.S.A.).
Exact type of the element damage was not investigated.
Images of homogeneous part of the phantom were acquired and stored in PACS archive (Sectra Imtec AB,
Linköping,...
Results
Images acquired with functional and damaged transducer were segmented at three different depths,
and minimum,
maximum,
median,
mode and mean pixel values were calculated across the image in the segments.
Image acquired with the functional transducer was homogeneous whereas the image acquired with damaged transducer showed several strong vertical streaks.
Calculated pixel values reflected visual analysis in functional transducer at all analyzed segments,
and median and mean pixel value showed least variation and noise across the image as expected.
Pixel values in image acquired with...
Conclusion
A novel method for the quality control of diagnostic ultrasound scanners was presented.
Median pixel value calculated from the superficial parts of an image is able to detect and locate damages to ultrasonic transducer.
It is simple and fast to perform,
requiring less than two minutes per transducer at the scanner,
and less than four minutes per transducer when registration of test patient is included.
Method provides intuitive parameter for monitoring the transducer throughout its lifespan.
Transducer deterioration can be monitored when the standard deviation...
References
[1] Weigang B,
Moore GW,
Gessert J,
Phillips WH,
Schafer M. The methods and effects of transducer degradation on image quality and the clinical efficacy of diagnostic sonography.
J Diagn Med Sonogr.
2003; 19(1):3–13
[2] Mårtensson M,
Olsson M,
Segall B,
Fraser AG,
Winter R,
Brodin L-Å. High incidence of defective ultrasound transducers in use in routine clinical practice.
Eur J Echocardiol.,
2009; 10(3):389–394
[3] Mårtensson M,
Olsson M,
Brodin L-Å. Ultrasound transducer function: annual testing is not sufficient.
Eur J Echocardiol.
2010; 11(9):801–805
[4]...
Personal Information
J.
V.
Aarnio,
J.
O.
Heikkinen.
Department of Medical Physics,
Mikkeli Central Hospital,
Porrassalmenkatu 35-37,
50100 Mikkeli,
Finland.
Mail to:
[email protected]
H.
Hermunen.
Department of Radiology,
Mikkeli Central Hospital,
Porrassalmenkatu 35-37,
50100 Mikkeli,
Finland