Purpose
Sentinel lymph node detection with iron oxide particles (Sentimag® technique) is a new technique. However, deposition of the iron oxide particles in the breast tissue may cause artifacts in the follow-up MRI examination (1). We aim to evaluate the effect of iron oxide particle deposition on follow-up MRI examination of patients who underwent sentinel lymph node detection with iron oxide particles.
Methods and materials
We have evaluated 217 patients who had sentinel lymph node detection with iron oxide particles. Follow-up MRI was performed in 35 of these patients (median age 49). The median time between surgery and MRI examination was 19 months (range: 3-36 months). Fat saturated T1W 3D gradient echo images without contrast and contrast enhanced images on the 120 sec after contrast administration were evaluated. The susceptibility artifact due to iron oxide particles were evaluated by two radiologists with experience in breast imaging in consensus. The susceptibility...
Results
Thirty-five patients were evaluated. The number of cases according to artifact grading were as follows: Grade 0: 11 (31.4%) Grade 1: 14 (40%) Grade: 2 (8.5%) and Grade 3: 7(20%).There was a statistical difference between segmental mastectomy and subcutaneous mastectomy (P=0.019) about the grade of the artifacts. Grade 3 artifacts were higher in breast conserving cases while Grade 0 was more frequent in subcutaneous mastectomy cases. Mean time span between the operation and MRI examination was 20 months. There was no correlation with the grade...
Conclusion
One third of the patients showed serious artifacts on both T1W and contrast enhanced images which obscured a proper evaluation. The artifacts were more prominent in segmental mastectomy cases but less seen on subcutaneous mastectomy cases (P=0.019). This is an expected outcome as in subcutaneous mastectomy cases the breast tissue is removed in a great extent and thin line of subcutaneous fat tissue remains. However, we have observed grade 1 artifacts in these cases which has a potential to still mask a millimetric local recurrence....
Personal information and conflict of interest
M. E. Aribal; Istanbul/TR - Speaker at GE Healthcare - Research/Grant Support at GE Healthcare C. Yilmaz; Istanbul/TR - nothing to disclose L. Çelik; Istanbul/TR - nothing to disclose C. Demirkiran; Istanbul/TR - nothing to disclose
References
Arslan G, Yılmaz C, Çelik L, Çubuk R, Tasalı N. Unexpected Finding on Mammography and MRI due to Accumulation of Iron Oxide Particles Used for Sentinel Lymph Nod Detection. Eur J Breast Health. 2019 Jul 1;15(3):200-202
Karakatsanis A, Obondo C, Abdsaleh S, Hersi AF, Eriksson S, Wärnberg F. Optimisation of breast MRI compatibility after sentinel node biopsy with paramagnetic tracers. Eur J Surg Oncol 2018; 44: 731-732. (PMID: 29478740)