Learning objectives
Recognise common pitfalls in ipsilateral localisation across craniocaudal (CC) and mediolateral oblique (MLO) DBT views that can lead to mis-targeted ultrasounds, missed subtle lesions and false-negative assessments.
Understand how dynamic AI-assisted tissue matching (auto-slice synchronisation with a confidence-scaled ellipse and uncertainty warnings) supports precise cross-view correlation, focusses workup efficiently and safely while supporting radiologists’ interpretation.
Apply a practical localisation checklist (warnings, coronal/equidistance references, and clock-face reporting) in daily practice to improve downstream ultrasound targeting and communication.
Background
BACKGROUND
Why localisation is hard in DBT: CC and MLO views are non‑orthogonal, tissue deforms between compressions and changes appearance, and each view contains ~60–80 depth slices; the reader must mentally map 3D tissue between views while scrolling. This increases cognitive load, fatigue and “false correspondence” (matching to the wrong tissue). [1,5–7]
Why it matters: incorrect cross‑view matching can lead to mis‑targeted ultrasound (US)/biopsy (false reassurance), unnecessary additional imaging (repeat targeted views or MRI), and reduced confidence/efficiency in assessment workflows. [1,2]
Why interval cancers matter:...
Findings and procedure details
C) Four cases: failure modes, imaging cues, and how to improveCase 1: Preventing a miss from a plausible manual mismatch: • Pitfall:“looks‑similar” posterior densities are manually paired across views, resulting in a false correspondence and US performed at the wrong site.•How AI helps:mapped tissue directs attention to the correct paired‑view location at index assessment (and flags uncertainty when geometry is unreliable).•Take‑home: treat easy appearance‑based matches with scepticism; triangulate using mapping + coronal/equidistance checks. [1]Case 2: Efficient US targeting when the correlate is occult in one...
Conclusion
DBT localisation errors are common because of non‑orthogonal views, tissue deformation and high slice burden. Dynamic AI‑assisted ipsilateral tissue matching with uncertainty warnings can reduce localisation errors—especially for non‑expert readers—and can be operationalised with a simple checklist to improve targeted US efficiency and safety while maintaining radiologist oversight. [1]
Personal information and conflict of interest
S. Morrell:
Research/Grant Support: King's College London
CEO: Elaitra Ltd
J. Cardoso:
Shareholder: Elaitra Ltd
B. Batohi:
Nothing to disclose
K. Satchithananda:
Advisory Board: Elaitra Ltd
R. K. Wasan:
Nothing to disclose
R. Rahim:
Nothing to disclose
S. Ourselin:
Advisory Board: Elaitra Ltd
E. F. Conant:
Advisory Board: Elaitra Ltd
J. C. Morel:
Advisory Board: Elaitra Ltd
References
[1] Morrell S, Hutel M, de Lucena O, et al. Dynamic AI‑assisted ipsilateral tissue matching for digital breast tomosynthesis.European Journal of Radiology.2025;112356. doi:10.1016/j.ejrad.2025.112356.[2] Gupta D, Friedewald SM. Lesion localization using digital breast tomosynthesis: where did I go wrong?Journal of Breast Imaging.2019;1(2):143‑150.[3] Dibden A, Offman J, Parmar D, et al. Reduction in interval cancer rates following the introduction of two‑view mammography in the UK breast screening programme.British Journal of Cancer.2014;110(3):560‑564.[4] Gilliland FD, Joste N, Stauber PM, et al. Biologic characteristics of interval and screen‑detected breast cancers.JNCI.2000;92(9):743‑749.[5]...