Purpose
Lumbar spine radiography is classified as a relatively high dose examination which irradiates the radiosensitive reproductive organs of both males and females.
Radiographic imaging of the lumbar spine accounts for 2.1% of all conventional X-ray examinations and 2.2% of the collective dose within the United Kingdom (UK) (Hart et al.,
2010).
One simple but effective method of radiation dose reduction is the replacement of the traditionally performed anteroposterior (AP) projection with the posteroanterior (PA) projection.Despite this it is still common practice for the majority of...
Methods and materials
Imaging equipment and phantom
The study was conducted in a university imaging department using a Wolverson Acroma X-Ray unit (Wolverson X-ray Ltd,
Willenhall,
UK) with a Varian 130 HS X-ray tube (Varian Medical Systems,
Palo Alto,
CA) with an inherent filtration of 3 mm aluminium. An Agfa (Agfa-Gevaert,
Mortsel,
Belgium) 35 x 43 cm computed radiography (CR) image receptor was used for acquisition and images were processed using an Agfa CR 35-X digitiser.
All exposures included the use of a 10:1 reciprocating grid with (40...
Results
Image quality
Total (weighted) image quality scores for both the AP and PA projections,
for each kVp increment,
are presented in Table 1and Figure 1.
An image quality score of 48 was considered equal to the reference image after weighting factors were applied.
Table 1. Image quality scores for the AP and PA projection.
kVp
AP Projection
PA Projection
Mean
SD
% change from reference
Mean
SD
% change from reference
75
41.6
1.1
-13.3
80
58.2
5.8
+21.3
45.4
5.8
-5.4
85
52.4
7.7...
Conclusion
Results obtained in this investigation demonstrate that the PA projection dramatically reduces the effective dose (up to 20%),
and absorbed doses to the stomach (70.4%),
colon (61.1%),
remainder tissues (33.2%),
ovaries (7.3%) and testes (15.9%),
when compared to the AP projection.
This may be at the expense of a minor reduction in image quality (not statistically significant).
This may further be considered acceptable when balanced against the significant dose reduction and,
therefore,
despite this the PA projection may be considered as the preferred option when...
Personal information
Enda Davey is a Band 5 Radiographer at Salford Royal Hospitals NHS Foundation Trust,
Manchester,
UK.
Enda was formerly a student at the University of Salford where this research was undertaken.
Andrew England (
[email protected]) is a Senior Lecturer at the University of Salford,
Manchester,
UK and was research supervisor for this project.
Research was conducted has part of the University of Salford Diagnostic Imaging Research Programme (DIRP).
References
European Commission.
European Guidelines on Quality Criteria for Diagnostic Radiographic Images.
Luxembourg: Commission of European Countries; 1996.
Fauber TL.
Radiographic imaging & exposure.
2nd ed.
ed.
St.
Louis,
Mo.: Mosby; 2004.
Hart D,
Wall BF,
Hiller MC,
Shrimpton PC.
Frequency and Collective Dose for Medical and Dental X-Ray Examinations in the UK,
2008.
In: Agency HP,
editor.
Oxford,
United Kingdom: Health Protection Agency; 2010.
Hogg P,
Blindell P.
Software for image quality evaluation using a force choice method. United Kingdom Radiological Congress; Manchester 2012.
p....