Description of activity and work performed
‘Is your Imaging EuroSafe?’ comprises a series of surveys on CT DRLs for different indications.
The 2016 survey campaign was to dedicated the collection of data to calculate DRLs for:
· Acute head trauma CT
· Acute stroke head CT
· Chest CT for pulmonary embolism
· Chest CT: rule out pulmonary metastasis
Results for acute head trauma CT
Representative CTDIvol and DLP values for acute head trauma CT |
|
n |
min |
max |
mean |
Q1 |
Q2 |
Q3 |
CTDIvol (mGy) |
22 |
22 |
82 |
48 |
38 |
46 |
57 |
DLP (mGy.cm) |
452 |
1501 |
844 |
658 |
793 |
968 |
|
|
|
|
|
|
|
|
Results for stroke head CT
Representative CTDIvol and DLP values for acute stroke head CT |
|
n |
min |
max |
mean |
Q1 |
Q2 |
Q3 |
CTDIvol (mGy) |
22 |
14 |
252 |
66 |
44 |
52 |
72 |
DLP (mGy.cm) |
206 |
3259 |
997 |
702 |
833 |
1114 |
Results for Chest CT: pulmonary embolism
Representative CTDIvol and DLP values for Chest CT: embolism |
|
min |
max |
mean |
Q1 |
Q2 |
Q3 |
CTDIvol (mGy) |
3 |
224 |
20 |
6,5 |
11 |
17,5 |
DLP (mGy.cm) |
91 |
1087 |
366 |
156 |
280 |
425 |
Results for Chest CT: rule out pulmonary metastasis
Representative CTDIvol and DLP values for Chest CT: metastasis |
|
min |
max |
mean |
Q1 |
Q2 |
Q3 |
CTDIvol (mGy) |
2 |
19 |
7 |
5 |
7 |
9,5 |
DLP (mGy.cm) |
67 |
876 |
318 |
181 |
252 |
431 |